Wednesday 8 April 2015

Analyze The Gas That Gives Us Life! - OXYGEN

       All of us know how essential is oxygen for our living…
As this is what happens without oxygen
                



Instrumentation engineers too realized the importance of oxygen thus they made an instrument “OXYGEN ANALYZER” for measuring the proportion of oxygen and displaying the oxygen content.

One among the primarily used oxygen analyzer is discussed

Zirconia Oxygen Analyzer (Direct Method):
This method gives the amount of oxygen using the concepts of electrochemistry. Here there is an electrolyte; 2 electrodes and a generated EMF (tells the oxygen concentration).
Let us have more insight of each component:
   1.     Zirconia sensor:
                                                  


The sensor probe uses a high-reliability zirconia sensor made up of Zirconium Dioxide, which is a special material; its resistance decreases drastically at higher temperatures (500-1000 deg C) and thus allows the conduction of ions as temperature increases. Using the property of zirconia that it conducts only oxygen ions at temperature of 600 deg C; zirconia acts as an electrolyte in the analyzer.

    2.       Reference and Measured Gas:
      The reference gas is usually taken as air which has a fixed amount of oxygen in it (20.9%) and measured gas is the one whose oxygen concentration has to be measured.



 
                                      

      
      3 .  Platinum electrodes:
It comprises of two electronically conducting, chemically inert electrodes attached to either side of a solid electrolyte tube. The plates provide a catalytic surface for the change in oxygen molecules.

       4.    Electric Oven, Thermocouple and Temperature Controller:
      In Zirconia sensor the temperature has to be above 4500C as only after that the zirconia become active as an electrolyte. Thus a thermocouple measures the temperature and controller gives the desired o/p to the electric heater and the desired temperature is maintained.
       
            Principle of Measurement
      There is a difference in concentration of oxygen (consequently a difference in the partial pressure of oxygen) on either side of the measured gas and reference gas which drives the oxygen to move from higher concentration to lower concentration.
                                           
Depending on the concentration on either side following reactions occur on each electrode:

                                                             
                                         
 


The oxygen ions formed flow through the electrolyte and the electrons through the outer wire as happens in the galvanic cell producing an EMF.
                            


Just a small review of class 8

A galvanic cell is an electrochemical cell which converts chemical energy to electrical energy through a spontaneous chemical reaction. In a galvanic cell there are two half-cells. Each half-cell contains an electrode in an electrolyte. The separation is necessary to prevent direct chemical contact of the oxidation and reduction reactions, creating a potential difference. The electrons released in the oxidation reaction travel through an external circuit before being used by the reduction reaction.




In oxygen analyzer we don’t need to have separate half cells as here zirconia is a solid acting as an electrolyte so no contact of the reaction takes place.

Let us move to high school and review some further concepts
The EMF generated by the galvanic cell in non-standard condition is given by the Nernst Equation:

                                


Where, E cell 0 is the cell potential under standard conditions (25oC, 1.0 M solution concentrations, 1.0 atm gas pressures), E cell  is the cell potential under nonstandard conditions, R is the gas constant 8.314 J/mol-K, T is the temperature in Kelvin, F is Faraday's constant (96,485 coulombs/mol), n is the number of electrons transferred in the balanced oxidation/reduction reaction, and Q is the reaction quotient ([products]/[reactants]).

Applying the Nernst Equation in Zirconia Oxygen Analyzer
The potential difference across the cell is given by Nernst equation:
                           


Here, E= E cell 0 -E cell all the terms represent the same terms as illustrated above and P1& P2 are the partial pressures of the oxygen on either side of the zirconia tube.
The Nernst equation can therefore be reduced to:
                          


Thus, if the oxygen partial pressure at one of the electrodes is known (reference gas-Air) and the temperature of the sensor is controlled (constant temperature for zirconia to act as a conducting electrolyte) and then measurement of the potential difference between the two electrodes enables the unknown partial pressure to be calculated using above equation. This partial pressure refers to the unknown oxygen concentration.
The following 3 cases can be there:
If P1>P2; ions flow from P1 to P2 i.e. a positive EMF
If P1<P2; ions flow from P2 to P1 i.e. a negative EMF
If P1=P2; no net ion flow i.e. a zero EMF
The analyzer electronically calculates the oxygen partial pressure, and therefore oxygen concentration, of a sample gas with unknown oxygen concentration. This is accomplished by measuring the potential, E, produced across the zirconium cell electrodes, substituting for E in the Nernst equation and anti-logging to obtain P1 (sample).

The zirconia oxygen analyzer may be used for the measurement of oxygen at any level between 
0-100% in gas mixtures.


But the Zirconia Oxygen Analyzer has a few drawbacks
1.       Incapable of measuring flammable gas mixtures due to the high temperature that needs to be maintained for the conductivity of zirconia.

2.       As there is a sensor that comes in direct contact with the reference gas, so can’t be used for corrosive 
      gases.
For overcoming these drawbacks another oxygen analyzer was introduced for knowing more about it stay tuned for the next article…  

No comments:

Post a Comment